Cytosensor Step-by-Step


 
Step 1: Cell Seeding
Step 2: Chamber Sensor Preparation
Step 3: Stabilization and Baseline Rate
Step 4: Test Material Dilution
Step 5: Dosing & Rate Data
Step 6: Data Analyses

Contents

Click on a number from above to reveal the corresponding step instructions. The links below are permanent and bookmarkable.

  1. Cell Seeding
  2. Chamber Sensor Preparation
  3. Stabilization and Baseline Rate
  4. Test Material Dilution
  5. Dosing & Rate Data
  6. Data Analyses

Cell Seeding

Cell Culture
Seeding of Cells

Stock cultures of L929 cells (mouse fibroblasts) are maintained at standard culture conditions until the cells reach the desired confluency.

When cells have reached the appropriate confluency, the cells are trypsinized, counted, and seeded into Transwell™ capsule cups. The seeded cups are incubated at standard culture conditions for at least 16 hours until use in the assay.

Chamber Sensor Preparation

Transfer of Transwell™ cell capsule
Cytosensor Chamber

The Transwell™ is placed into the cytosensor chamber after at least 16 hours of incubation.

Stabilization and Baseline Rate

Transfer of Chamber
Stabilization

The chambers are transferred to the Cytosensor Microphysiometer. The L929 cells are ready for stabilization.

During the stabilization period, rates of cellular metabolism are displayed every minute for at least one hour.

At the end of the hour, at least 5 control rates are determined and averaged for each chamber. This average value is used as the baseline for the individual chamber.

Test Material Dilution

Cytosensor Serial Dilution

A serial dilution is prepared for each test article, with each dilution series covering a range of at least 7 concentrations.

Test materials must be aqueous soluble to be tested in this assay system.

Dosing & Rate Data

Dosing and Rate Data

After baseline rates are determined for all of the active chambers, a 20 minute dosing cycle begins.

In this cycle, each test material concentration (dose) is added to the chambers for a 13.5 minute period. The cycle continues with a 6-minute rinse with a Low-Buffered DMEM to remove the dose. After the rinse cycle is complete, an acidification (metabolic) rate is determined within 30 seconds.

Data Analyses

Cytosensor Data

Metabolic rates after each test material concentration (dose) are measured and compared to the baseline values. A percent of control metabolic rate is determined for each dose, and that value is plotted on a dose response curve to determine a MRD50 value.